19 Light scattering is not limited to a single particle. Particles are moving in solution, bouncing into and off of each other, and light does the same. Since we are concerned with elastic scattering, this phenomenon does not result in a loss of intensity, but it becomes more challenging to accurately measure particle diffusion if the light path is interrupted by more than one particle on its way to the detector. DLS instrumentations generally work with very small sample volumes. This allows for fewer multi-scattering events because there are fewer particles in the light's pathway, so it is more likely the light will only be scattered by a single particle. Additionally, instrumentation has very precisely focused lasers and detectors in order to keep the area being measured small. DLS instruments are carefully designed to minimize the effects of multi-scattering events in solution. Multi-scattering events